C120 Dual-Flow BioAerosol Sampler
Product Features
4.3’’ capacitive touchscreen,rapid response
All components and circuits interfaces are isolated from the external environment ro prevent oxidant intrusion
Long Battery Life
100L/min≥10hours (60m³)
200L/min≥4hours (48m³)
Flow Accuracy
±2.5%(100L/min)
±3% (200L/min)
Compliance with FDA21 CFR Part 11
Remote Sampling
Powerful and stabilized pump to support Remote/Isolator sampling(up to 8m)
*The remote/isolator sampling can only support the flow rate of 100L/min*
Operation Interface
Data
- Intuitive sampling records
- Search function for quick search of records - One-click for exporting and printing
Sampling
- Intelligent sampling mode enables continuous and cyclic sampling
-The sampling process is clear and distinct
LED lighting and beep prompt error
Setup
- Smart Settings mode
- User-friendly input method
Specification
FLOW RATE: 100L/min 200L/min
FLOW RATE ACCURACY: ±2.5%
ZONE: Over 20000 locations
SAMPLING VOLUMES: 1-6000L, 6 default volumes
DELAY: 0-59minutes59seconds
CYCLES: 1-99cycles
INTERVAL: 0-59minutes59seconds
SAMPLING HEAD: Φ0.85mm*280holes
DIMENSIONS: Φ110mm, H: 202mm (without foot mat)
WEIGHT: 2.7kg(6.0lbs)
EXTERNAL SURFACE: Anodized aluminium
POWER: 24V±10% AC adapter
BATTERY: 10.8V, 6800mAh rechargeable lithium battery, running for over 10 hours
DATA STORAGE: 20000 sets of sample records
DISPLAY: 4.3" capacitive touch screen
LANGUAGE: Chinese, English
CALIBRATION FREQUENCY: Recommend once a year
OPERATING CONDITIONS TEMPERATURE:5℃-30℃, relative humidity up to 80%
STORAGE CONDITIONS TEMPERATURE: 0℃-40℃, relative humidity up to 80%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 61000-6-1:2007, EN 61000-6-3:2007+A1
WARRANTY: 2 years after activation
Warranty
We offer a two-year warranty for the DBAS including hardware and software from the date of activation. It is considered activation when continues sampling for 40 minutes or sampling for 10 times. The repair service and parts replacement are warranted for 6 months from the date of the shipment.
In these cases, the warranty is void.
a. The warranty is void if the DBAS is repaired and disassembled by the company or person without authorization.
b. Damage is due to mishandling or operator misuse.
We offer the free remote upgrade for lifetime when the users require during use.
The DBAS is a precision instrument. It must be operated by trained laboratory personnel only.