Thermal Printer Particle Counter
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Operation Interface
Data
-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports
-Data review and data export
1. Sum up data by zone.
2.Freely switch counting unit between △/Σ/#,m /ft.
3.Automatically generate UCL values and average values
4.Search for sample records precisely by time,zone and operator.
-Data integrity: featured with audit trail.
Sampling
- Real-time concentration curves, to get a better understanding of environmental particle condition.
- Quick editing function, to make sampling more flexible.
Setup
- Clear layout for orders. Simple and easy to use.
- One-click for system software upgrade.
- Multiple sets of personalized sampling schemes
can be customized for any zone.
Sampling
Delay/ Interval | In auto mode, the first sample of each location complies with the delay. The following samples comply with the interval. In the case of manual mode, all samples comply with the delay. The sampling can be stopped during delay and interval. Press the start icon again to continue the sampling. In the case of custom mode, delay and interval can be changed at any moment. The countdown mode is used for delay and interval. When the sampling interval time is more than or equal to 2min, the blower stops running after the first cycle ends for 10s, and the blower starts running again at 20s before the next cycle. |
Volume | This shows the current sampled volume (above the line) and total volume of the sample (below the line). In NORM mode, the volume cannot be changed. |
Sample Time | This shows the current sampled time (above the line) and total time of the sample (below the line). In NORM mode, the sample time cannot be changed. |
Samples | This shows the number of samples which are being taken in the current location (left of the split) and the total number of the samples (right of the split). |
Auto Sample | If checked, the sampling will stop after all the samples on one location are finished. |
Auto Print | If checked, the data will be printed automatically after each sampling is complete. If NORM mode is selected, data of one location will be printed when sampling(s) of this location are all completed. If manual mode is selected, data of each sample will be printed when this sample is completed. |
Graph | The curves show the instantaneous data of the 0.5μm and 5.0μm particle channel in real time. |
Location display | It shows the number of locations in the sampling room. Flashing green light indicates the sampling is underway in this location. Green light indicates the location which the sampling has completed. If alarm appears during the sampling, this location should appear in red light. |
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records
DISPLAY: 8” LCD capacitive touch screen
LANGUAGE: Chinese, English
DATA RELIABILITY: Compliant with 21 CFR Part 11
REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std
CALIBRATION FREQUENCY: Recommend at least once a year
DIMENSIONS (HXWXD): 254x204x263mm
WEIGHT: 6.1Kg (13.45lbs) (without battery)
ENCLOSURE: 316L Stainless Steel
POWER: AC 100-240V, 50 Hz/60 Hz
BATTERY: 10.8V, 9000mAh, rechargeable lithium battery
BATTERY ENDURANCE: A110: running for 4 hours
A120: running for 7 hours
A130: running for 9 hours
MAX POWER: consumption AC 90W
OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%
STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B
WARRANTY: 2 years after activation