Portable Particle Counter
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Operation Interface
Data
-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports
-Data review and data export
1. Sum up data by zone.
2.Freely switch counting unit between △/Σ/#,m /ft.
3.Automatically generate UCL values and average values
4.Search for sample records precisely by time,zone and operator.
-Data integrity: featured with audit trail.
Sampling
- Real-time concentration curves, to get a better understanding of environmental particle condition.
- Quick editing function, to make sampling more flexible.
Setup
- Clear layout for orders. Simple and easy to use.
- One-click for system software upgrade.
- Multiple sets of personalized sampling schemes
can be customized for any zone.
Sampling
Description
Group
Touch【Group】icon to open the group list, and select the group in which the room to sample is.
Zone
Touch【Zone】icon to open the room list, and select the sampling room.
Sampling SCM
Touch【Sampling SCM】icon to open the sampling scheme list, and select the sampling SCM. One room can be associated with several schemes. And one scheme can be associated with several rooms.
Sampling Rule
Sampling rule consists of NORM mode and Custom mode. The NORM mode is used for the samplings per ISO/EUGMP/CHINESEGMP/Fed Std. Under this mode, parameters cannot be modified after the room and sampling scheme are selected (the background of sampling conditions turn to grey.) If the sampling process is stopped, the result of the standard report is invalid. The Custom mode is used for sampling for other purposes. Under this mode, sampling parameters such as location, volume, sample time can be modified and a sampling can be stopped at any moment.
Locations
The location cannot be changed under the NORM mode. It can be a selection list option under the Custom mode.
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records
DISPLAY: 8” LCD capacitive touch screen
LANGUAGE: Chinese, English
DATA RELIABILITY: Compliant with 21 CFR Part 11
REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std
CALIBRATION FREQUENCY: Recommend at least once a year
DIMENSIONS (HXWXD): 254x204x263mm
WEIGHT: 6.1Kg (13.45lbs) (without battery)
ENCLOSURE: 316L Stainless Steel
POWER: AC 100-240V, 50 Hz/60 Hz
BATTERY: 10.8V, 9000mAh, rechargeable lithium battery
BATTERY ENDURANCE: A110: running for 4 hours
A120: running for 7 hours
A130: running for 9 hours
MAX POWER: consumption AC 90W
OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%
STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B
WARRANTY: 2 years after activation