WIFI Airborne Particle Counter for Environmental
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Operation Interface
Data
-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports
-Data review and data export
1. Sum up data by zone.
2.Freely switch counting unit between △/Σ/#,m /ft.
3.Automatically generate UCL values and average values
4.Search for sample records precisely by time,zone and operator.
-Data integrity: featured with audit trail.
Sampling
- Real-time concentration curves, to get a better understanding of environmental particle condition.
- Quick editing function, to make sampling more flexible.
Setup
- Clear layout for orders. Simple and easy to use.
- One-click for system software upgrade.
- Multiple sets of personalized sampling schemes
can be customized for any zone.
Admin
Admin is used to set account policy, create and modify accounts. Only the ADMIN account has the right to enter this interface for relevant settings.
Title | Description |
Account Policy | Touch【】icon to open the setting menu of account policy, and set the password expiration date, password length and number of login attempts. |
Maximum Password Age | Password never expires: the password is permanently valid. Expires in days: after the number of days set, the password will be prompted password expired when logging in the account, and the password needs to be modified. The password needs to be changed compulsively in the login interface, otherwise the device cannot be logged in. |
Minimum Password Age | Allow changes immediately: the password can be changed at any time. Expires in days: the password cannot be changed within the number of days set. |
Minimum Password Length | Permit blank password: Allow the save password to be null. At least characters: the password length needs to reach the set number of characters. |
Unsuccessful Log-on attempts | Disable: No limit to the number of login attempts. Enable: When log-on, the number of unsuccessful log-on attempts reach the set number of times, the account is disabled. The ADMIN account is not limited to the number of the log-on attempts. |
Account | The default account is ADMIN, and the password is 0. When creating a new account, enter the account name. |
Group | The level of account is classified as admin, supervisor and operator. |
Enabled | ON: indicates the account is enabled. OFF: indicates that the account is disabled. |
Access Authority | The operator account can be configured for different access authority. Access authority include zone, sampling SCM, data SCM, display, time, info alert, device, custom sampling, WIFI, MV connect, audit trail, offline print, export data, and clear samples. |
Password | Set and confirm the password, the two passwords must be the same. The password can be saved as null. |