VHP-resistant Particle Counter
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Sampling
Touch to display the set channel threshold. | |
Set the counting mode, counting unit and size channel for the data displaying during the sampling. Counting mode consists of ∑ cumulative count and Δ differential count. Cumulative count: the number of particles in and above the selected channel. Differential count: the number of particles between the selected channel and adjacent channel. Counting unite: #, #/ft3, #/m3. Size channel: 0.5/1.0/2.0/3.0/5.0/10.0µm (A110) or 0.3/0.5/1.0/3.0/5.0/10.0µm (A120, A130). All of the particle size channel or part of the particle size channel can be selected to display | |
Print the sample data. | |
If there is an abnormal situation during the latest sampling process, the sampling result may be deviated from the actual sampling environment, and a new sampling can be conducted through this function. Touch【】icon, depending on the sampling condition of the lastest sample, a new sampling is conducted. |
When all the sampling conditions are set, touch【
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records
DISPLAY: 8” LCD capacitive touch screen
LANGUAGE: Chinese, English
DATA RELIABILITY: Compliant with 21 CFR Part 11
REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std
CALIBRATION FREQUENCY: Recommend at least once a year
DIMENSIONS (HXWXD): 254x204x263mm
WEIGHT: 6.1Kg (13.45lbs) (without battery)
ENCLOSURE: 316L Stainless Steel
POWER: AC 100-240V, 50 Hz/60 Hz
BATTERY: 10.8V, 9000mAh, rechargeable lithium battery
BATTERY ENDURANCE: A110: running for 4 hours
A120: running for 7 hours
A130: running for 9 hours
MAX POWER: consumption AC 90W
OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%
STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B
WARRANTY: 2 years after activation