Waterproof Airborne Particle Counter for Compressed Air
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Operation Interface
Data
-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports
-Data review and data export
1. Sum up data by zone.
2.Freely switch counting unit between △/Σ/#,m /ft.
3.Automatically generate UCL values and average values
4.Search for sample records precisely by time,zone and operator.
-Data integrity: featured with audit trail.
Sampling
- Real-time concentration curves, to get a better understanding of environmental particle condition.
- Quick editing function, to make sampling more flexible.
Setup
- Clear layout for orders. Simple and easy to use.
- One-click for system software upgrade.
- Multiple sets of personalized sampling schemes
can be customized for any zone.
Sampling SCM
Sampling SCM is used for setting up the scheme for sampling. The scheme consists of volume, sample time, delay, etc. It can be added, amended or removed in this menu. The sampling SCM can be associated with zone here. Touch【Sampling】icon of the left menu in the setting interface to set the scheme.
Description
Name
The name of sampling SCM (up to 19 characters). Add: touch the input area to enter a new scheme, touch【】icon to save the settings. Touch【】icon to select an existing scheme from the pull-down menu. Touch【】icon to find a scheme, and touch【OK】icon to confirm. Bulk sampling scheme addition: select the existing sampling scheme from the pull-down menu, modify the name of the scheme and touch【】icon to save, can select between【Cover the original sampling SCM】icon or【Save as a new sampling SCM】icon. The scheme “default” is a system default sampling scheme. It can be modified, but it cannot be deleted. After modifying the name of the Default scheme, touch【】icon to save it as a new scheme.
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records