PRODUCT FEATURE | |
8 inch capacitive touchscreen, quick response Even with two pairs of latex gloves | Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water |
Compliance with FDA21 CFR Part 11 | |
AUTOMATICALLY COLLECT AIR SAMPLES | |
APC data exceed the limit,triggers the BAS to start sampling | |
OPERATION INTERFACE | |
|
1. Sum up data by zone. 2. Freely switch counting unit between △/Σ/#, m /ft. 3. Automatically generate UCL values and average values 4. Search for sample records precisely by time, zone and operator.
|
| |
|
|
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft⊃3;at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records
DISPLAY: 8” LCD capacitive touch screen
LANGUAGE: Chinese, English
DATA RELIABILITY: Compliant with 21 CFR Part 11
REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std
CALIBRATION FREQUENCY: Recommend at least once a year
DIMENSIONS (HXWXD): 254x204x263mm
WEIGHT: 6.1Kg (13.45lbs) (without battery)
ENCLOSURE: 316L Stainless Steel
POWER: AC 100-240V, 50 Hz/60 Hz
BATTERY: 10.8V, 9000mAh, rechargeable lithium battery
BATTERY ENDURANCE: A110: running for 4 hours
MAX POWER: consumption AC 90W
OPERATING CONDITIONS TEMPERATURE: 5℃-35℃, relative humidity up to 95%
STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B
WARRANTY: 2 years after activation
VIDEO